Home › MCQs › Laboratory Operations › Instrumentation & methods

Laboratory Operations: Instrumentation & methods – page 2

39 Laboratory Operations MCQs on Instrumentation & methods with answers and explanations.

12
Q21MediumInstrumentation & methods

During serum protein electrophoresis at pH 8.6, gamma globulins move slightly toward the cathode even though they carry a small negative charge. This is caused by:

Answer: D. Electroendosmosis

Electroendosmosis is the bulk flow of buffer toward the cathode caused by fixed negative charges on the support medium. Weakly charged gamma globulins are carried with it.

ID MG-LOP-0014 · Found a mistake? Report it
Q22MediumInstrumentation & methods

In reversed-phase HPLC with a nonpolar C18 column and a polar mobile phase, which compounds elute first?

Answer: A. The most polar compounds

In reversed-phase chromatography the stationary phase is nonpolar, so polar compounds interact least and elute first. Nonpolar compounds are retained longer.

ID MG-LOP-0015 · Found a mistake? Report it
Q23MediumInstrumentation & methods

In a time-of-flight (TOF) mass analyzer, such as in MALDI-TOF identification of bacteria, which ions reach the detector first?

Answer: D. Ions with lower mass-to-charge ratio

All ions receive similar kinetic energy, so lighter ions (lower m/z) travel faster through the flight tube and arrive first. Neutral molecules are not accelerated or detected.

ID MG-LOP-0017 · Found a mistake? Report it
Q24MediumInstrumentation & methods

In real-time PCR, sample A has a cycle threshold (Ct) of 22 and sample B has a Ct of 32 for the same target. What does this indicate?

Answer: D. Sample A contains much more starting target than sample B

Ct is inversely related to the starting amount of target: fewer cycles are needed to cross the threshold when more template is present. A 10-cycle difference is roughly a 1000-fold difference.

ID MG-LOP-0019 · Found a mistake? Report it
Q25MediumInstrumentation & methods

A 100 µL pipette is checked gravimetrically with water (assume density 1.000 g/mL). The mean delivered mass is 0.0980 g. The inaccuracy (percent error) is:

Answer: A. −2.0%

Delivered volume = 98.0 µL. Error = (98.0 − 100)/100 × 100 = −2.0%. The gravimetric method converts the weight of delivered water to volume.

ID MG-LOP-0020 · Found a mistake? Report it
Q26MediumInstrumentation & methods

Most clinical osmometers measure osmolality by detecting which colligative property of the sample?

Answer: C. Decrease in freezing point

Freezing point depression osmometers measure how much solutes lower the freezing point (1 osmol/kg lowers it by 1.86 °C). Refractive index is not a colligative property.

ID MG-LOP-0021 · Found a mistake? Report it
Q27MediumInstrumentation & methods

In atomic absorption spectrophotometry for lead, the light source is a hollow cathode lamp whose cathode is made of:

Answer: B. Lead

The cathode is made of the same element being measured so it emits that element's exact wavelengths, which ground-state atoms in the flame absorb.

ID MG-LOP-0022 · Found a mistake? Report it
Q28MediumInstrumentation & methods

In an ion-selective electrode analyser, the potassium electrode membrane usually contains:

Answer: B. Valinomycin

Valinomycin is an ionophore that binds K⁺ selectively and is used in potassium ISE membranes. Sodium is measured with a special sodium-selective glass membrane.

ID MG-LOP-0109 · Found a mistake? Report it
Q29MediumInstrumentation & methods

In a Severinghaus pCO₂ electrode, carbon dioxide is measured by detecting:

Answer: B. A pH change in a bicarbonate solution behind a gas-permeable membrane

CO₂ diffuses through the membrane into bicarbonate solution, forming carbonic acid and lowering pH, which a glass electrode senses (potentiometric). Current at a platinum cathode describes the Clark pO₂ electrode.

ID MG-LOP-0110 · Found a mistake? Report it
Q30MediumInstrumentation & methods

The Clark electrode used to measure pO₂ works on which principle?

Answer: C. Amperometry: current from reduction of oxygen

The Clark electrode applies a fixed voltage; oxygen is reduced at the platinum cathode and the current is proportional to pO₂. Potentiometric glass electrodes measure pH and pCO₂.

ID MG-LOP-0111 · Found a mistake? Report it
Q31MediumInstrumentation & methods

The main advantage of a double-beam spectrophotometer over a single-beam instrument is that it:

Answer: B. Corrects for fluctuations in lamp output and detector response

By splitting light between sample and reference paths and comparing them, a double-beam instrument cancels variations in source intensity. It still needs a monochromator and appropriate blanks.

ID MG-LOP-0114 · Found a mistake? Report it
Q32MediumInstrumentation & methods

Isoelectric focusing separates proteins according to their:

Answer: D. Isoelectric point in a pH gradient

In isoelectric focusing each protein migrates in a pH gradient until it reaches the pH where its net charge is zero (pI), then stops. SDS-PAGE, not IEF, separates mainly by size.

ID MG-LOP-0116 · Found a mistake? Report it
Q33MediumInstrumentation & methods

A glass pipette marked 'TD' is calibrated:

Answer: D. To deliver the stated volume

TD (to deliver) pipettes deliver the marked volume when drained; TC (to contain) pipettes hold the marked volume and must be rinsed out.

ID MG-ELOP-0035 · Found a mistake? Report it
Q34MediumInstrumentation & methods

Which light source is commonly used in spectrophotometers for measurements in the ultraviolet range?

Answer: C. Deuterium lamp

Deuterium lamps give a continuous UV spectrum; tungsten lamps are used for the visible range, and hollow cathode lamps for atomic absorption.

ID MG-ELOP-0038 · Found a mistake? Report it
Q35MediumInstrumentation & methods

Which grade of laboratory water, with the highest purity, is used for sensitive tests such as trace analysis and molecular methods?

Answer: A. Type I (clinical laboratory reagent water)

Type I (CLRW) water has the highest resistivity and lowest contamination; Type III is used for rinsing glassware.

ID MG-ELOP-0040 · Found a mistake? Report it
Q36HardInstrumentation & methods

At high absorbance values, a spectrophotometer's standard curve bends toward the concentration axis and loses linearity. The most likely instrumental cause is:

Answer: C. Stray light reaching the detector

Stray light adds unabsorbed light at the detector, making absorbance read falsely low; the effect is largest in highly absorbing samples, causing negative deviation from Beer's law.

ID MG-LOP-0009 · Found a mistake? Report it
Q37HardInstrumentation & methods

A patient with severe hypertriglyceridemia has sodium 126 mmol/L by indirect ISE but 140 mmol/L by direct ISE on a blood gas analyzer. The best explanation is:

Answer: D. Pseudohyponatremia due to the electrolyte exclusion effect in the indirect method

Indirect ISE dilutes the sample and assumes normal plasma water; excess lipid lowers plasma water, giving falsely low sodium. Direct ISE measures activity in plasma water and is unaffected.

ID MG-LOP-0023 · Found a mistake? Report it
Q38HardInstrumentation & methods

A chloridometer generates silver ions at a constant rate; the time to reach the endpoint is proportional to the chloride concentration. This method is:

Answer: D. Coulometric-amperometric titration

In coulometric-amperometric titration, generated Ag⁺ combines with Cl⁻; excess Ag⁺ at the endpoint increases current and stops the timer. ISE methods measure a voltage instead.

ID MG-LOP-0112 · Found a mistake? Report it
Q39HardInstrumentation & methods

A patient on peritoneal dialysis with icodextrin has high glucose meter readings but a normal plasma glucose on the laboratory analyser. Which meter enzyme is most likely used?

Answer: B. Glucose dehydrogenase with PQQ

GDH-PQQ also reacts with maltose, which is formed from icodextrin, giving falsely high readings. Glucose oxidase meters are specific for glucose and not affected by maltose.

ID MG-LOP-0118 · Found a mistake? Report it
12