Home › MCQs › Hematology › Hematology methods & instruments

Hematology: Hematology methods & instruments

83 Hematology MCQs on Hematology methods & instruments with answers and explanations.

12345
Q1EasyHematology methods & instruments

For a manual leukocyte count, 20 µL of blood is added to 380 µL of diluting fluid. What dilution factor is used in the calculation?

Answer: A. 20

20 µL in a final volume of 400 µL gives a 1:20 dilution, the standard for manual WBC counts. A 1:200 dilution is used for manual RBC counts.

ID LG-HEM-0027 · Found a mistake? Report it
Q2EasyHematology methods & instruments

A marrow aspirate is needed from a 20-year-old adult. Which site is preferred?

Answer: C. Posterior iliac crest

In adults, red marrow persists mainly in the axial skeleton and proximal long bones; the posterior superior iliac crest is safe and cellular. The tibia is used only in infants.

ID LG-HEM-0029 · Found a mistake? Report it
Q3EasyHematology methods & instruments

Before Romanowsky staining, a thin blood film should be fixed with which reagent?

Answer: D. Absolute methanol

Methanol fixes cells without lysing them. Water-based or acidic reagents such as acetic acid would lyse red cells; acid alcohol is a decolorizer.

ID LG-HEM-0030 · Found a mistake? Report it
Q4EasyHematology methods & instruments

Which microscope part collects light from the source and focuses it onto the specimen?

Answer: A. Condenser

The substage condenser concentrates and directs light through the specimen. The objective and ocular magnify the image.

ID LG-HEM-0093 · Found a mistake? Report it
Q5EasyHematology methods & instruments

The anticoagulant abbreviated EDTA stands for what?

Answer: C. Ethylenediaminetetraacetic acid

EDTA (ethylenediaminetetraacetic acid), usually as the K2 salt, is the anticoagulant of choice for CBCs because it preserves cell morphology.

ID LG-HEM-0189 · Found a mistake? Report it
Q6EasyHematology methods & instruments

Romanowsky stains combine methylene blue/azure dyes with eosin. Which stain does NOT belong to this group?

Answer: D. Harris hematoxylin

Giemsa, Wright and Leishman are Romanowsky stains producing the purple nuclear effect. Hematoxylin is a mordant nuclear dye used in H&E and Pap staining.

ID LG-HEM-0214 · Found a mistake? Report it
Q7EasyHematology methods & instruments

Which procedure is a bleeding-time test rather than a method for measuring hemoglobin?

Answer: B. Ivy method

Ivy's method measures bleeding time from a standardized forearm incision. The other three convert hemoglobin to a colored derivative read photometrically or visually.

ID LG-HEM-0253 · Found a mistake? Report it
Q8EasyHematology methods & instruments

An elevated ESR is most consistent with:

Answer: B. Active inflammation with raised fibrinogen

Acute-phase proteins such as fibrinogen and immunoglobulins promote rouleaux, raising the ESR. High red cell counts, low globulins and abnormal red cell shapes lower it.

ID LG-HEM-0273 · Found a mistake? Report it
Q9EasyHematology methods & instruments

Which total leukocyte count lies within the usual adult reference interval?

Answer: D. 7.5 × 10^9/L

The adult WBC reference interval is roughly 4.0–11.0 × 10^9/L. Lower values indicate leukopenia and higher values leukocytosis.

ID LG-HEM-0300 · Found a mistake? Report it
Q10EasyHematology methods & instruments

Thin blood films are routinely fixed before Romanowsky (e.g. Giemsa) staining with:

Answer: A. Absolute methanol

Methanol fixes cells without lysing red cells or distorting morphology. Heat fixation is used for bacterial smears; formalin and mercuric chloride are histology fixatives.

ID LG-BAC-0198 · Found a mistake? Report it
Q11EasyHematology methods & instruments

Reticulocytes are counted after staining living (unfixed) red cells. Which dye is used for this supravital staining?

Answer: C. New methylene blue

New methylene blue (or brilliant cresyl blue) precipitates residual RNA in living reticulocytes. Wright's, Giemsa and Field's are Romanowsky stains used on fixed smears.

ID LG-HIS-0267 · Found a mistake? Report it
Q12EasyHematology methods & instruments

In the electrical impedance (Coulter) principle, the size of each voltage pulse is proportional to:

Answer: B. The volume of the cell

A cell passing through the aperture displaces conductive diluent and raises resistance; the pulse height reflects cell volume and the number of pulses gives the count. Granularity is assessed by side scatter, not impedance.

ID MG-HEM-0046 · Found a mistake? Report it
Q13EasyHematology methods & instruments

In flow cytometry, forward scatter and side scatter mainly indicate, respectively:

Answer: A. Cell size and internal granularity or complexity

Light scattered at low angles (forward) relates to cell size, while light scattered at 90 degrees (side) relates to granules and internal structure. Antigen density is measured by fluorescence.

ID MG-HEM-0049 · Found a mistake? Report it
Q14EasyHematology methods & instruments

The cyanmethemoglobin (hemiglobincyanide) reference method for hemoglobin measures absorbance at:

Answer: C. 540 nm

Hemiglobincyanide has a broad absorbance peak at 540 nm, which is used in the ICSH reference method. 340 nm is used for NADH/NADPH enzyme assays.

ID MG-HEM-0050 · Found a mistake? Report it
Q15EasyHematology methods & instruments

When reading a microhematocrit, including the buffy coat in the red cell column will:

Answer: B. Falsely increase the hematocrit

Only the packed red cell column should be read; the buffy coat of white cells and platelets adds to the reading. A falsely high Hct would lower the calculated MCHC, not raise it.

ID MG-HEM-0059 · Found a mistake? Report it
Q16EasyHematology methods & instruments

In the Westergren ESR method (CLSI/ICSH), blood is typically:

Answer: C. Diluted 4:1 with sodium citrate and read after 60 minutes

The classic Westergren method uses 4 parts blood to 1 part 3.8% (or 3.2%) sodium citrate in a 200 mm tube, read at 60 minutes. Undiluted EDTA blood is used only in modified methods.

ID MG-HEM-0061 · Found a mistake? Report it
Q17EasyHematology methods & instruments

In a flow cytometer, which component converts emitted fluorescent light into an electronic signal?

Answer: B. Photomultiplier tube

Photomultiplier tubes (or photodiodes) detect light and convert it into electrical pulses for analysis. Dichroic mirrors only direct light of certain wavelengths to the correct detector.

ID MG-HEM-0070 · Found a mistake? Report it
Q18EasyHematology methods & instruments

When counting cells in a hemacytometer square, cells that touch the boundary lines should be:

Answer: C. Counted on two adjacent sides only (e.g., top and left)

To avoid counting the same cell twice in neighboring squares, cells touching two chosen sides are included and those touching the opposite two are excluded. Counting all four sides gives falsely high results.

ID MG-HEM-0323 · Found a mistake? Report it
Q19EasyHematology methods & instruments

Automated reticulocyte counting identifies reticulocytes by:

Answer: A. Staining residual RNA with a nucleic-acid dye and measuring the signal

Fluorescent or absorbing dyes bind RNA, and the amount of signal separates reticulocytes from mature cells and grades their maturity (IRF). Impedance alone cannot detect RNA.

ID MG-HEM-0328 · Found a mistake? Report it
Q20EasyHematology methods & instruments

The blue laser used in most clinical flow cytometers to excite FITC and PE emits light at about:

Answer: B. 488 nm

A 488 nm blue laser excites FITC and PE and is standard in clinical instruments. A 633–640 nm red laser is used for dyes such as APC.

ID MG-HEM-0332 · Found a mistake? Report it
12345