Home › MCQs › Laboratory Operations › Quality control & QA

Laboratory Operations: Quality control & QA

54 Laboratory Operations MCQs on Quality control & QA with answers and explanations.

123
Q1EasyQuality control & QA

A single control result outside ±3 SD violates the Westgard 1-3s rule and most often signals:

Answer: A. Random error; reject the run

1-3s is a rejection rule mainly sensitive to random error (e.g. pipetting, bubbles). Patient results must not be released until the cause is found and corrected.

ID MG-LOP-0001 · Found a mistake? Report it
Q2EasyQuality control & QA

A control has a mean of 200 mg/dL and an SD of 5 mg/dL. The coefficient of variation is:

Answer: D. 2.5%

CV = SD ÷ mean × 100 = 5 ÷ 200 × 100 = 2.5%. CV lets you compare the precision of methods at different concentrations.

ID MG-LOP-0003 · Found a mistake? Report it
Q3EasyQuality control & QA

Out of 100 people with a disease, a test is positive in 90. The test's sensitivity is:

Answer: A. 90%

Sensitivity = true positives ÷ (true positives + false negatives) = 90 ÷ 100 = 90%. Specificity would need results from people without the disease.

ID MG-LOP-0004 · Found a mistake? Report it
Q4EasyQuality control & QA

In the classic Westgard multirule procedure, the 1-2s rule is used as:

Answer: A. A warning rule that triggers inspection with other rules

In the classic multirule scheme, 1-2s is a warning that prompts checking the other rules. Using it alone for rejection causes a high false-rejection rate (about 5% per control).

ID MG-LOP-0087 · Found a mistake? Report it
Q5EasyQuality control & QA

Repeated control results are tightly grouped but far from the established target value. The method is best described as:

Answer: D. Precise but inaccurate

Close agreement between replicates means good precision; distance from the true value means poor accuracy (bias). Precision and accuracy are independent characteristics.

ID MG-LOP-0090 · Found a mistake? Report it
Q6EasyQuality control & QA

A potassium control has a mean of 4.0 mmol/L and an SD of 0.1 mmol/L. The ±2 SD acceptable range is:

Answer: C. 3.8–4.2 mmol/L

±2 SD = 4.0 ± (2 × 0.1) = 3.8 to 4.2 mmol/L, which includes about 95% of expected values. 3.7–4.3 is the ±3 SD range.

ID MG-LOP-0092 · Found a mistake? Report it
Q7EasyQuality control & QA

The main purpose of proficiency testing (external quality assessment) is to:

Answer: B. Compare a lab's accuracy with other labs using unknown samples

Proficiency testing sends samples of unknown value to many labs and grades each result against peers or target values, assessing accuracy externally. Daily precision is monitored by internal QC.

ID MG-LOP-0094 · Found a mistake? Report it
Q8EasyQuality control & QA

A reference interval is usually defined as the central 95% of results from healthy people. Therefore, among healthy individuals:

Answer: D. About 5% will have results outside the interval

The central 95% excludes 2.5% at each end, so about 1 in 20 healthy people will have an 'abnormal' result. 32% applies to a ±1 SD range.

ID MG-LOP-0101 · Found a mistake? Report it
Q9EasyQuality control & QA

A glucose result exceeds the upper limit of the analytical measurement range. The appropriate action is to:

Answer: A. Dilute the sample by a validated protocol and reanalyse

Results outside the AMR are unverified. The sample should be diluted using a validated dilution and the result multiplied by the dilution factor, or reported as greater than the limit per policy.

ID MG-LOP-0106 · Found a mistake? Report it
Q10EasyQuality control & QA

A new test is performed on 200 people known to be free of the disease. It is negative in 170 and positive in 30. What is its specificity?

Answer: C. 85%

Specificity = true negatives / all people without disease = 170/200 = 85%. The false-positive rate is 30/200 = 15%.

ID MG-LOP-0152 · Found a mistake? Report it
Q11EasyQuality control & QA

A specimen is labelled with the wrong patient's name at the bedside. This error occurs in which phase of testing?

Answer: C. Pre-analytical

Patient identification and specimen labelling happen before analysis, so it is a pre-analytical error, the phase where most laboratory errors occur. Analytical errors happen during measurement.

ID MG-LOP-0153 · Found a mistake? Report it
Q12EasyQuality control & QA

A correct result is transcribed into the wrong patient's record during manual entry. This is an error in which phase?

Answer: D. Post-analytical

Reporting, transcription and communication of results are post-analytical steps. The measurement itself was correct, so it is not an analytical error.

ID MG-LOP-0154 · Found a mistake? Report it
Q13EasyQuality control & QA

A potassium result of 6.9 mmol/L is a critical value. The best practice when reporting it by telephone is to:

Answer: C. Report promptly to a responsible caregiver and ask for a read-back

Critical values need prompt communication to a responsible licensed caregiver, with read-back of the result and documentation of who received it. Waiting for the clinician to see it in the system may delay urgent care.

ID MG-LOP-0155 · Found a mistake? Report it
Q14EasyQuality control & QA

Which international standard specifies requirements for quality and competence of medical laboratories?

Answer: D. ISO 15189

ISO 15189 is the specific standard for medical laboratory quality and competence used for accreditation. ISO 14001 covers environmental management, and ISO 9001 is a general quality standard.

ID MG-LOP-0163 · Found a mistake? Report it
Q15EasyQuality control & QA

Precision of a laboratory method refers to:

Answer: A. How close repeated measurements of the same sample are to each other

Precision is reproducibility; closeness to the true value is accuracy.

ID MG-ELOP-0009 · Found a mistake? Report it
Q16EasyQuality control & QA

In a normal (Gaussian) distribution of control results, about what percentage of values lie within ±2 SD of the mean?

Answer: C. 95%

About 68% of values lie within ±1 SD, 95% within ±2 SD and 99.7% within ±3 SD.

ID MG-ELOP-0010 · Found a mistake? Report it
Q17EasyQuality control & QA

Specificity of a diagnostic test is the proportion of:

Answer: A. People without the disease who test negative

Specificity = true negatives ÷ (true negatives + false positives); the proportion of diseased people testing positive is sensitivity.

ID MG-ELOP-0012 · Found a mistake? Report it
Q18EasyQuality control & QA

Most laboratory errors occur in which phase of testing?

Answer: B. Pre-analytical phase

Errors in ordering, patient identification, collection, labelling and transport account for most laboratory errors.

ID MG-ELOP-0013 · Found a mistake? Report it
Q19EasyQuality control & QA

A written, step-by-step document describing exactly how a test is performed in the laboratory is called a:

Answer: A. Standard operating procedure (SOP)

SOPs ensure every staff member performs the test the same way and are part of document control.

ID MG-ELOP-0016 · Found a mistake? Report it
Q20MediumQuality control & QA

Seven consecutive control values fall on the same side of the mean, all within ±2 SD. This is a:

Answer: B. Shift, suggesting systematic error

A sudden run of values on one side of the mean is a shift, often from a new reagent lot or calibration change. A gradual drift in one direction is a trend (e.g. lamp ageing).

ID MG-LOP-0002 · Found a mistake? Report it
123
📱 Practise with a timer, track your score and earn certificates in the free MLT Globe app – Google Play or practise online.